YCIT-J Series Medical IT Insulation Monitoring Instrument

YCIT-J Series Medical IT Insulation Monitoring Instrument

YCIT-J Series Medical IT Insulation Monitoring Instrument

The YCIT-J Series Medical IT Insulation Monitor is a high-precision safety device designed for medical isolated power systems. Powered by a 32-bit ARM processor and a 2.8-inch TFT color touchscreen, it provides real-time monitoring of insulation resistance, transformer temperature, and electrical parameters (voltage, current, and harmonics). Featuring audible and visual alarms, it ensures the highest level of electrical safety for both patients and medical staff by detecting faults before they become hazards.

YCIT-J Series Medical IT Insulation Monitoring Instrument

The YCIT-J Series Medical IT Insulation Monitor uses a high-performance 32-bit ARM processor as its data processing core. It is equipped with a color TFT touchscreen as the human-machine interface, providing users with an intuitive and friendly interaction experience.

Designed specifically for medical isolated power systems, this device monitors IT circuits downstream of the medical isolation transformer in real-time. It displays critical parameters such as insulation resistance, load current, and transformer temperature. When monitored values exceed preset limits, the device triggers audible and visual alarms, allowing medical and technical staff to troubleshoot faults promptly and significantly reducing the risk of electric shock for both patients and healthcare personnel.

Key Features

  • Insulation Monitoring: Monitors insulation resistance in both AC and DC isolated power systems.
  • Electrical Parameter Measurement: Tracks voltage, current, apparent power, and active power.
  • Temperature Monitoring: Monitors the temperature of the isolation transformer (primary and secondary).
  • Dual Power Input Monitoring: Monitors voltage for both mains and backup power supplies.
  • Advanced Diagnostics: Monitors residual current, output waveforms, and harmonic content (to provide a basis for harmonic management).
  • Alarm System: Audible and visual alarms for insulation faults, transformer over-temperature, and system overload/residual current.
  • Self-Testing: Includes a built-in self-test and calibration function.
  • Communication: Equipped with 2-way RS485 communication ports.
  • Display: 2.8-inch TFT color touchscreen.

ItemTechnical Specifications
Insulation ResistanceRange: 30–5000 kΩ; Relative Error: ±10%; Test Voltage: <12V; Alarm Setting: 30–5000 kΩ (adjustable)
Voltage MeasurementRange: 30–500 Vac; Accuracy: RMS 0.2%; Alarm Setting: 50–300 Vac (adjustable)
Current MeasurementRange: 0.5–120A (includes residual current); Accuracy: RMS 0.5%; Resolution: 0.005A; Alarm Setting: 5–100A (adjustable)
Power (Apparent)Accuracy: 0.5%; Resolution: 0.001 kVA; Alarm Setting: 1–20 kVA (adjustable)
Power (Active)Accuracy: 0.5%; Resolution: 0.001 kW
TemperatureRange: -30°C to +210°C; Accuracy: ±2°C; Alarm Setting: 20°C to 200°C (adjustable)
HarmonicsMeasurement Range: 0–22nd harmonic
RS485 CommunicationType: Two-wire half-duplex; Baud Rate: 600–4800 bps; Protocol: Custom Enterprise; Distance: 2000m
Working VoltageDC 12V (Supplied by dedicated company power source)
EnvironmentOperating Temp: -10°C to +40°C; Limit Temp: -15°C to +60°C; Humidity: ≤95% (Non-condensing)
Display ModeLCD / Color TFT Touchscreen
InstallationEmbedded (Panel Mount) or DIN Rail Mount
Altitude≤ 2000m